Accessories:
Pressure Regulator
Installation Kit
Maintenance Kit
Accuracy:
± 5 % or ±0.03 mg/L (ppm) as CL₂, whichever is greater. *For permanganate accuracy, please refer to permanganate application notes
Air Purge:
0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting
Alarm:
Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Certifications:
Europe, CE Approved with:
EN 61326-1
CISPR 11
EN 61010-1
IEC 60529
North America:
UL 61010A-1
CAN/CSA C22.2 No. 1010.1-92
Communication Capabilities:
4-20mA Output
Cycle Time:
2.5 min
Description :
CL17 Chlorine Analyzer with Pressure Regulator and Reagents for Total Chlorine
Dimensions (H x W x D):
454 mm x 314 mm x 179 mm
Display:
LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line
Drain:
1/2-inch ID flexible hose, hose barb
Enclosure Construction:
ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Inlet:
1/4-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure:
1 to 5 psig (0.07 to 0.34 bar), 1.5 psi is optimum
Interferences:
Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr⁶⁺ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃.
Light source:
Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
Lower Limit of Detection (LOD):
0.03 mg/L (ppm)
Mounting:
Wall mount
Operating Humidity:
Up to 90 % at 40 °C (104 °F) maximum
Operating temperature range:
5 - 40 °C (41 - 104 °F)
Parameter:
Chlorine, Free & Total
Pressure Range:
Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Pressure Regulator Included?:
Yes
Range:
0 - 5 mg/L Cl₂
Recorder Outputs:
One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground.
Repeatability:
0.01 ppm or 3.0 % , whichever is greater
Sample Flow Rate:
200 - 500 mL per minute minimum
Sample Temperature:
5 - 40°C (41 - 104°F)
Warranty:
12 months
Weight:
10.49 kg
What's included?:
Hach CL17 Chlorine Analyzer, a one-month supply of reagents, 100 mL of Sulfuric acid, Maintenance kit, Installation kit with pressure regulators, and Manual. (The power cord is ordered separately.)