Chemistries                 0800 50 55 66

9187 sc Chlorine Dioxide amperometric analyser

9187 sc Chlorine Dioxide amperometric analyser
Product #: LXV434.99.00001
Unit Price Contact Hach

Includes a selective membrane which avoids interference from Chlorine

Stable measurements thanks to the flow-through cell

Easily installed premounted flow-through system

Minimal maintenance thanks to automatic cleaning

Compatible with multi-sensor, multi-parameter digital controllers

What's included?

Hach 9187sc Chlorine Dioxide sensor panel, 4 spare membranes, and Manual.

Specifications

Accuracy: 5% or ±10 ppb ClO2, whichever is greater
Calibration Interval: 2 months
Calibration Method: Zero Calibration: Electrically or with ozone-free water
Calibration: Comparison of lab method with process sample
Enclosure Rating: IP66 / NEMA 4X
Flow: 14 L/h (200 - 250 mL/min) auto-regulated by flow through cell
Height : 270 mm
Interferences: None
Lower Limit of Detection (LOD): 0.01 mg/L ClO2
Maintenance Interval: 6 months for membrane and electrolyte (measurement cell)
Measuring Interval: Continuous
Measuring Principle: Amperometric/Membrane (electrode,
membrane, electrolyte)
Mounting: Flat, vertical surface (panel, stand, etc.)
Operating temperature range: 0 - 45 °C
Parameter: Chlorine Dioxide
Pressure Range: 0.1 - 2 bar in flow cell
Range: 0 - 2 mg/L ClO2
Relative Humidity: 0 - 90% non-condensing
Response time: 90% in less than 90 seconds
Sample Input: 0.25 " outer diameter
Sample Output: 0.5 " inner diameter
Sample Temperature: 2 - 45 °C
Standard deviation: 1.5 %
Storage conditions: -20 - 60 °C
Temperature Compensation: Automatic over sample temperature range
  Automatic over sample temperature range
Warranty: 12 months
Weight: 6.5 kg
What's included?: Hach 9187sc Chlorine Dioxide sensor panel, 4 spare membranes, and Manual.
Width: 250 mm