Accuracy:
Turbidity: Defined according to ISO/WD 13530. Suspended Solids of reading or ±0.001 NTU, whichever is greater
Accuracy 2:
Defined according to ISO/WD 13530.Turbidity Suspended Solids Less than 5 % of reading (depends on homogeneity of municipal activated sludge)
Cable Length:
10 m (optional extension cables available)
Calibration:
Turbidity Suspended Solids:Formazin or StablCal® Standard
Based on gravimetric TSS analysis with a correction factor procedure
Certifications:
CE certified to EN 61326-1, EN 61326/A1, EN 61326/A2, EN 61010-1
Construction - Sensor Body (RD 242/262):
Wiper
Sensor body: Silicon
Stainless steel
Detection Limit:
Turbidity
Suspended Solids 0.001 mg/L
Diameter Sensor:
(diameter x length)
Insertion sensor: 60 x(2.4 x)
Flow:
Flow Velocity 3 m/s (9.8 ft./s) maximum
Length:
(diameter x length) (12.4 in.) Insertion sensor:
Lower Limit of Detection (LOD):
Turbidity
Suspended Solids 0.001 NTU
Material:
Optics Carrier and Sleeve: stainless steel 1.4571 or black PVC
Wiper Arm: stainless steel 1.4581; Wiper Blade: silicone (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571
Threaded cable fitting: stainless steel 1.4305 or white PVC
Stainless steel can be susceptible to corrosion at chloride concentrations ≥ 500 mg/l.
Max. cable length Sensor-to-Controller:
100 m
Measuring Range Turbidity:
Suspended Solids 0.001 - 4000 NTU
Mounting:
Fixed to the rim of the tank
Through sidewall of a pipeline using a ball valve; minimum pipe size (4 in.) in carbon or stainless steel
Operating temperature range:
0 - 40 °C
Parameter:
obs
Pressure Range:
Sample Pressure 6 bar (87 psi) maximum
Range:
Turbidity
Suspended Solids 0.001 mg/L 50 g/L
Repeatability:
Defined according to ISO/WD 13530.Turbidity Suspended Solids Less than3% of reading (depends on the homogeneity of municipal activated sludge)
Repeatability Turbidity:
Defined according to ISO/WD 13530. Suspended Solids. Less than1% of reading
Response time:
Initial response in 1 second
Sample Temperature:
0 - 40°C (32 - 104°F)
Signal Average Time:
Initial response in 1 second
Turbidity:
Units of Measure Turbidity
Suspended Solids user selectable—NTU, FNU, or TE/F
Units:
Units of Measure
Turbidity
Suspended Solids: User selectable—g/L, mg/L, ppm, or % solids
What's included?:
Turbidity & Suspended Solids probe, controller, adapter, flange, user manual