Accuracy:
5% or ±10 ppb ClO2, whichever is greater
Calibration Interval:
2 months
Calibration Method:
Zero Calibration: Electrically or with ozone-free water
Calibration: Comparison of lab method with process sample
Enclosure Rating:
IP66 / NEMA 4X
Flow:
14 L/h (200 - 250 mL/min) auto-regulated by flow through cell
Height :
270 mm
Interferences:
None
Lower Limit of Detection (LOD):
0.01 mg/L ClO2
Maintenance Interval:
6 months for membrane and electrolyte (measurement cell)
Measuring Interval:
Continuous
Measuring Principle:
Amperometric/Membrane (electrode,
membrane, electrolyte)
Mounting:
Flat, vertical surface (panel, stand, etc.)
Operating temperature range:
0 - 45 °C
Parameter:
Chlorine Dioxide
Pressure Range:
0.1 - 2 bar in flow cell
Range:
0 - 2 mg/L ClO2
Relative Humidity:
0 - 90% non-condensing
Response time:
90% in less than 90 seconds
Sample Input:
0.25 " outer diameter
Sample Output:
0.5 " inner diameter
Sample Temperature:
2 - 45 °C
Standard deviation:
1.5 %
Storage conditions:
-20 - 60 °C
Temperature Compensation:
Automatic over sample temperature range
Automatic over sample temperature range
Warranty:
12 months
Weight:
6.5 kg
What's included?:
Hach 9187sc Chlorine Dioxide sensor panel, 4 spare membranes, and Manual.
Width:
250 mm