Accuracy:
5% of measured value +0.2 mg/L (with standard solutions) NH4-N and NO3-N
Cable Length:
Standard: 10 m (33.8 ft)
Extension cables are available as an option in the following lengths:
5, 10, 15, 20, 30, 50 m (16.4, 33.8, 49.2, 65.6, 98.4, 164 ft)
Maximum overall length: 100 m (328 ft)
Calibration Method:
With Cartrical plus technology: automatic import of factory calibration data from cartridge to probe by RFID; 1 and 2 point matrix correction
Controller Compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Data Transfer:
Current output, relays and bus communication via SC controller
Dimensions (D x L):
84.5 mm x 320 mm
Flow:
< 4 m/s max. velocity
Includes:
Ammonium & Nitrate probe, sensor cartridge, user manual
Installation Angle:
45 ° ± 15 ° (vertical in flow direction)
Length:
320 mm
Lower Limit of Detection (LOD):
0.2 mg/L NH4–N and NO3-N
Material:
Cartridge: stainless steel (1.4571), PVC, POM, ABS, NBR
Cleaning unit: TPE, PUR, stainless steel (1.4571)
Sensor: stainless steel (1.4571), ASA + PC, silicon, PVC and PU
Measurement method:
Potentiometric ion-selective electrodes for ammonium, potassium, nitrate and chloride, reference system and temperature sensor
Measuring Interval:
Continuous
Measuring uncertainty:
± 5 % + 0.2 mg/L (with standard solution) NH4-N and NO3-N
Operating temperature range:
Air: -20 - 45 °C (-4 - 113 °F)
Parameter:
NH4-N and NO3-N
pH Range:
5 - 9 pH
Process Connection:
1 inch thread
Protection Class:
IP68
Range:
0 - 1000 mg/L NH₄-N
0 - 1000 mg/L NO₃-N
0 - 1000 mg/L K+
0 - 1000 mg/L Cl-
Repeatability:
± 5 % of the measured value (with standard solution)+ 0.2 mg/L NH4-N and NO3-N
Response time:
< 3 min
Sample Pressure:
0.3 bar max.
Sample Temperature:
2 - 40 °C (35.6 - 104 °F)
Sensor Cartridge:
With Cartrical plus technology: compact housing containing calibrated electrodes for ammonium, potassium, nitrate and chloride, reference system and temperature sensor, all calibrated to each other; typical lifetime 12 months
Sensor Immersion Depth:
0.3 - 3.0 m
max. 0.3 bar (4.4 psi) pressure
Storage conditions:
5 to 40 °C (41 to 104 °F) sensor cartridge
Warranty:
12 months
Weight:
2.38 kg
What's included?:
Ammonium & Nitrate probe, sensor cartridge, user manual