Prices Online Include GST      0800 50 55 66

CLT10sc Total Chlorine Analyzer

CLT10sc Total Chlorine Analyzer
Product #: LXV45B.99.11022
Unit Price Contact Hach

Compatible with Hach's "Plug and Play" Digital Controllers

The CLF10 sc and CLT10 sc Chlorine Analyzers are designed to be used with all Hach sc digital controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.

No Reagent Replacement, No Waste Stream

Chlorine measurement with the CLF10sc and CLT10sc amperometric technology eliminates the need for routine reagent replacement and waste stream management.

EPA Compliance

The CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements in accordance with EPA Method 334.0. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required for Method 334.0.

Advanced Diagnostics

CLF10sc and CLT10sc advanced diagnostics alert you when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting sample flow sensor for notification of insufficient sample flow.

Real-time Process Monitoring and Control

The CLF10sc and CLT10sc Chlorine Analyzers allow for real-time monitoring and control of disinfection processes.

What's included?

Hach CL10sc Panel, 1 M digital extension cable, Panel Manual, Chlorine Sensor Manual.

Specifications

Accuracy: Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2: Free Chlorine: ± of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3: Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4: Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection: 5 pin, M12 connector
Cable Length: 1 m (between gateway and sc controller)
Calibration Method: 1-point or 2-point (zero and slope) calibration
Certifications: CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line: 1/2 in ID
Connections: 1/4 in OD
Controller: Panel only
Controller Compatibility: sc controller platform
Description : CLT10sc Total Chlorine Analyzer
Dimensions: mm x 25 mm (7.68 in x 0.98 in)
  Sensor Only: 195
Fitting type: USA
Includes: CL10sc Panel, 1 m digital extension cable, Panel Manual, Chlorine Sensor Manual.
Interferences: Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length: Sensor: 195 mm
Lower Limit of Detection (LOD): 30 ppb (0.03 ppm)
Material: Corrosion-resistant materials, fully-submersible
(stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement method: Reagentless, electrochemical, three-electrode amperometric system
Mounting: Panel or Wall Mounting
Operating temperature range: 0 - 45 °C
Options: Panel Only
Parameter: Total Chlorine
pH Monitoring Required?: None
pH Range: 4 - 9
Power Requirements (Voltage): None
Pressure Range: 0.5 bar, no pressure impulses and/or vibrations
Range: 0 - 20 ppm Cl₂
Repeatability: 30 ppb or 3%, whichever is greater
Response time: Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
Response Time: Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate: 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature: 5 - 45°C (41 - 113°F)
Storage conditions: -20 °C to 60 °C dry
Warranty: 12 months
  1 Year
Weight: Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?: Hach CL10sc Panel, 1 M digital extension cable, Panel Manual, Chlorine Sensor Manual.