Chemistries                 0800 50 55 66

EZ1037 Sulphide Analyser

EZ1037 Sulphide Analyser
Product #: EZ1037.XXXXXXXX
Unit Price Contact Hach

What's included?

EZ1037 Sulphide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Reagent Glass Containers with Fittings (for Colour, Buffer)

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatic cleaning: Yes
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / ETL certified
Cycle Time: 15 min (dilution + 5 min.)
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Strong reducing agents like Hydrosulphide, Nitrite > 0,2 mg/L, Sulphite > 200 mg/L and Thiosulphate > 5 mg/L interfere by reducing the blue colour. Oxygen reacts with Sulphide and must be avoided. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 50 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: galvanised steel, powder coated
Measurement method: Colorimetric measurement at 670 nm using N,N-dimethyl-p-phenylenediamine dihydrochloride (NNDP), conform with standard method ASTM 4500-S2- D
Model: EZ1037
Number of sample streams: 1, 2, 4, or 8
Parameter: Sulphide
Power: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Power supply: 100 - 240 VAC, 50/60 Hz
Precision: Better than 2% full scale range for standard test solutions
Protection Class: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Range: 50 - 500 µg/L
  50 - 500 µg/L S2-

Optional:
0.2 - 2 mg/L (with internal dilution)
0.4 - 4 mg/L (with internal dilution)
5 - 50 mg/L (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 1 year
  12 months
Weight: 25 kg
What's included?: EZ1037 Sulphide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Reagent Glass Containers with Fittings (for Colour, Buffer)