Alarm:
                                            
                                            
                                        
									    1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
                                    
                                
    						
								
                                    
                                        
                                            
                                                Ambient Temperature:
                                            
                                            
                                        
									    10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Analog Outputs:
                                            
                                            
                                        
									    Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Automatic cleaning:
                                            
                                            
                                        
									    Yes
                                    
                                
    						
								
                                    
                                        
                                            
                                                Calibration:
                                            
                                            
                                        
									    Automatic, 2-point; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Certifications:
                                            
                                            
                                        
									    CE compliant / UL certified
                                    
                                
    						
								
                                    
                                        
                                            
                                                Cooling water:
                                            
                                            
                                        
									    Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
                                    
                                
    						
								
                                    
                                        
                                            
                                                Cycle Time:
                                            
                                            
                                        
									    30 min including digestion of 10 min (standard)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Demineralised water:
                                            
                                            
                                        
									    For rinsing
                                    
                                
    						
								
                                    
                                        
                                            
                                                Digital outputs:
                                            
                                            
                                        
									    Optional: Modbus (TCP/IP, RS485)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Dimensions (H x W x D):
                                            
                                            
                                        
									    690 mm x 465 mm x 330 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Drain:
                                            
                                            
                                        
									    Atmospheric pressure, vented, min. Ø 32 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Earth connection:
                                            
                                            
                                        
									    Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
                                    
                                
    						
								
                                    
                                        
                                            
                                                Instrument air:
                                            
                                            
                                        
									    Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interferences:
                                            
                                            
                                        
									    Antimony (III), Bismuth (III), Chloroplatinate, Gold (III), Iron (III), Lead (II), Mercury (II), Metavanadate, and Silver (I) can precipitate with Nitrate. The presence of Copper (II) may decompose the diazonium salt which results in a low result. Strong oxidising agents. NCl₃ results in a false red colour. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Lower Limit of Detection (LOD):
                                            
                                            
                                        
									    ≤ 250 µg/L
                                    
                                
    						
								
                                    
                                        
                                            
                                                Material:
                                            
                                            
                                        
									    Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
                                    
                                
    						
								
                                    
                                        
                                            
                                                Measurement method:
                                            
                                            
                                        
									    Colorimetric measurement at 546 nm using hydrazine reduction and NEDD colour solution after persulphate digestion in alkaline medium, conform with APHA 4500-N
                                    
                                
    						
								
                                    
                                        
                                            
                                                Model:
                                            
                                            
                                        
									    EZ7702
                                    
                                
    						
								
                                    
                                        
                                            
                                                Number of sample streams:
                                            
                                            
                                        
									    1, 2, 4, or 8
                                    
                                
    						
								
                                    
                                        
                                            
                                                Parameter:
                                            
                                            
                                        
									    Total Nitrogen (TN)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Power:
                                            
                                            
                                        
									    230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
                                    
                                
    						
								
                                    
                                        
                                            
                                                Precision:
                                            
                                            
                                        
									    Better than 3% full scale range for standard test solutions
                                    
                                
    						
								
                                    
                                        
                                            
                                                Protection Class:
                                            
                                            
                                        
									    Analyser cabinet: IP55 / Panel PC: IP65
                                    
                                
    						
								
                                    
                                        
                                            
                                                Range:
                                            
                                            
                                        
									    0.25 - 10 mg/L TN
                                    
                                
    						
								
                                    
                                        
                                            
                                                Reagent Requirements:
                                            
                                            
                                        
									    Keep between 10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample Flow Rate:
                                            
                                            
                                        
									    100 - 300 mL/min
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample Pressure:
                                            
                                            
                                        
									    By external overflow vessel
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample Quality:
                                            
                                            
                                        
									    Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample Temperature:
                                            
                                            
                                        
									    10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Validation:
                                            
                                            
                                        
									    Automatic; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Warranty:
                                            
                                            
                                        
									    12 months
                                    
                                
    						
								
                                    
                                        
                                            
                                                Weight:
                                            
                                            
                                        
									    25 kg
                                    
                                
    						
								
                                    
                                        
                                            
                                                What's included?:
                                            
                                            
                                        
									    EZ7702 Total Nitrogen Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x empty 2.5L Reagent Container with Fittings (Buffer Solution), 2 x empty 5L Reagent Containers with Fittings (Oxidation B & Reductor Solution), 1 x empty 10L Reagent Container with Fittings (Oxidation A Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (Colour Solution)