Accuracy:
0.01 - 2 ppb: ± 0.1 ppb
2 - 10,000 ppb: ± 5%
Acidity:
< 50 ppm
Ambient Temperature:
5 - 50 °C
Analog Outputs:
6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum
Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) recommended, twisted pair shielded wire
Calibration Method:
Automatic with known addition
Manual: 1 or 2 points
Dimensions:
681 mm x 452 mm x 254 mm (H x W x D)
Display:
Coloured 5.7” LCD
Electrode Type:
Sodium ISE (ion specific electrode) electrode and reference electrode with KCl electrolyte
Fuse:
Input power: T 1.6 A, 250 VAC
Relays: T 5.0 A, 250 V
Include Autocalibration?:
No
Include Cation Kit?:
No
Inlet:
Sample line and sample bypass drain: 6 mm O.D. push-to-connect fitting for plastic tubing
Chemical and case drains: 7/16 inch I.D. slip-on fitting for soft plastic tubing
Interference Phosphate 10 ppm:
< 0.1 ppb
Load of Analog Outputs:
600 Ohm
Lower Limit of Detection (LOD):
0.01 ppb
Maintenance Interval:
Every 90 days: refill electrolyte, reactivation, conditioning, and calibration solution
Material:
Polyol case, PC door, PC hinges and latches, 304/316 SST hardware
Max. Concentration of Suspended Solids in Sample:
< 2 NTU, no oil, no grease
For boiler sample type install approx. 100 µm filter
Mounting:
Panel mount
Number of analog outputs:
6
Number of Channels:
4
Number of relays:
6
Options:
Analyser only
Parameter:
Sodium
pH Range:
6 - 10 pH
Pollution Degree:
2
Power requirements (Hz):
50/60 Hz
Power Requirements (Voltage):
100 - 240 VAC
Protection rating:
IP65, PCBA housing
Range:
0.01 ppb - 10,000 ppb
Relative Humidity:
10 - 80%, non-condensing
Relay output:
6; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Connection: 1.0 - 1.29 mm² (18 - 16 AWG) wire; 1.0 mm² (18 AWG) stranded recommended, 5 - 8 mm O.D. cable
Repeatability:
< 0.02 ppb or 1.5% reading (whichever is greater) within ± 10 °C variation
Response time:
From 0.1 ppb to 10 ppb: T90 ≤ 3 minutes, T95 ≤ 4 minutes
From < 1 ppb to 100 ppb: T90 < 2 minutes, T95 < 3 minutes (about 150 s)
Sample conditioner:
Di-isopropylamine (DIPA) (1 L/100 days) at 25 °C for a sample pH target of 10.5
Sample Flow Rate:
100 - 150 mL/min (6 - 9 L/h)
Sample Pressure:
0.2 - 6 bar
Sample Temperature:
5 - 45 °C
Storage conditions:
-20 - 60 °C
Warranty:
12 months
Weight:
14 kg with empty bottles
What's included?:
Hach NA5600sc Sodium Analyser, 4 channel, with 4 channel installation kit and user manuals, reference electrode sodium, sodium ion selective electrode, empty DIPA bottle, the cover for DIPA bottle and the tray for DIPA bottle which is installed in the analyser.