Accuracy:
(Sample temp. 20 - 50°C within ± 5°C of calibration temperature)
The greater of
Membrane 29561A: ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar
Membrane 2952A: ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar
Membrane 2935A: ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar
Accuracy 2:
(Sample temp. 0 - 50°C independent of calibration temperature)
The greater of
Membrane 2952A: ±3% of reading, or ±60 ppb, or ±0.6 cc/kg, or ±20 mbar
Membrane 2935A: ±3% of reading, or ±150 ppb, or ±2.5 cc/kg, or ±50 mbar
Application:
Power
Certifications:
2004/108/EC - EN 61326-1
Cycle Time:
17 s
Flow Rate:
Typical Flow Rate
Membrane 29561A: 220 mL/min
Membrane 2952A: 200 mL/min
Membrane 2935A: 100 mL/min
Parameter:
Hydrogen (H₂)
Protection Class:
IP65
Purge connection:
Nitrogen Purge
Radiation Resistance Level:
Membrane 29561A: 105 rad
Membrane 2952A: 108 rad
Membrane 2935A: 108 rad
Range:
Range at 25 °C
Membrane 29561A: 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar
Membrane 2952A: 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar
Membrane 2935A: 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar
Response Time:
17 s
Sample Pressure:
0 - 20 bar
Sample Temperature:
0 - 50 °C (32 - 122 °F)
Temperature Range:
-5 - 50 °C
Warranty:
6 months
Weight:
0.95 kg
What's included?:
Sensor only.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.