Prices Online Include GST      0800 50 55 66

Solitax highline sc Turbidity (0.001-4000 NTU) and Suspended Solids (0.001-500 g/L) built-in probe, w/o wiper, stainless steel

Solitax highline sc Turbidity (0.001-4000 NTU) and Suspended Solids (0.001-500 g/L) built-in probe, w/o wiper, stainless steel
Product #: LXV424.99.02200
Unit Price Contact Hach

Accurate, Color-independent Measurement

Primary and active sludge with different structures and colors, dark digested sludge and light lime sludge make the highest demands on the precise measurement of solids. Only SOLITAX sc solids probes with the patented dual scattered light method can satisfy these demands.

Easy One-point Calibration

All SOLITAX sc probes are factory calibrated in conformity with DIN EN ISO 7027 for long-term calibration stability. Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. Multipoint calibration is available for nonlinear samples.

Excellent correlation to laboratory analysis

Comparisons between laboratory analysis and continuous suspended solids measurement using SOLITAX sc sensors show an exceptional correlation to mixed liquor, settled and thickened sludge. Online measurement not only saves time on manual analysis but also provides critical real time TSS measurements that can be used to operate the plant more efficiently.

Improve overall efficiency of the plant

Connects directly to all Hach sc controllers for immediate use. Helps improve process control and reduce treatment costs associated with polymer use, digester volume, and sludge handling.

Improve overall efficiency of the plant

Connects directly to all Hach sc controllers for immediate use. Helps improve process control and reduce treatment costs associated with polymer use, digester volume, and sludge handling.

What's included?

Turbidity/Suspended Solids probe, user manual

Specifications

Accuracy: Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper: No
Body Material: Stainless Steel
Cable Length: 10 m (optional extension cables available)
Calibration Method: Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications: CE
Controller: Sensor only
Diameter Sensor: 60 mm
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Length: 315 mm
Maintenance Interval: 1 h/month
Material: Wiper: Silicon
Sensor body: Stainless steel 1.4571
Max. cable length Sensor-to-Controller: 100 m
Measuring Principle: Infrared Duo scattered light technique for color-independent
turbidity measurement Turbidity in accordance with
DIN EN ISO 7027 / TS equivalent DIN 38414
Model: Solitax highline sc
Mounting: Insertion
Operating temperature range: 0 - 40 °C
Parameter: High Range Suspended Solids,
Turbidity
Pressure Range: Stainless steel insertion sensor: <6 bar or <60 m
Range: TSS-content: 0.001 - 500 g/L; 0.001 - 500,000 mg/L (the upper range is dependent on the sample property)
Turbidity: 0.001 - 4000 NTU
Repeatability: TSS content: < 3 %
Turbidity: < 1 %
Response Time T90: 1 - 300 s adjustable
Special Notes: Installation note:
Distance sensor-wall
> 50 cm (Turbidity)
Warranty: 12 months
Weight: Insertion stainless steel: 2.4 kg (5.29 lb.)
What's included?: Turbidity/Suspended Solids probe, user manual