Accuracy:
±5% of reading or ±0.1 NTU (whichever is greater) from 0 - 2000 NTU
Accuracy 2:
±10% of reading from 2000 - 9999 NTU
Alarm:
Relays can be programmed through sc controller for warning, alarm, feeder, event, pulse width modulation, and frequency control.
Ambient Temperature:
0 - 50 °C
Channels:
Instrument Only
Controller:
No
Enclosure Construction:
Plastic instrument enclosure
Enclosure Rating:
NEMA 12
Installation category:
Category II
Mounting:
Wall mount
Number of Channels:
None
Power Requirements (Voltage):
100/230 VAC
Range:
0.1 - 9999 NTU
Relative Humidity:
5 - 95% non-condensing
Repeatability:
1.0% or ±0.04 NTU, whichever is greater
Resolution:
<1000 NTU: 0.01 NTU
1000 - 9999.9 NTU: 0.1 NTU
Response time:
Initial response in 45 seconds
Sample Flow Rate:
Sample flow required 1.0 - 2.0 L/min
Sample Input:
0.75 " NPT female thread
Sample Output:
1 " NPT female thread
Sample Temperature:
0 - 50 °C
0 - 50°C (32 - 122°F)
Storage conditions:
-20 - 80 °C
Warranty:
24 months
Weight:
15.8 kg (34.8 lb)
What's included?:
SS7 sc, Wall Mount Kit, Cylinder Brush, Light Source Alignment Template, 0.75 x 0.75 inch NPT Nipple, Calibration Cup, 0.75 inch NPT Ball Valve, SS7 sc User Manual, and 500 mL 4000 NTU Formazin Standard.