Accuracy:
± 5% or ± 0.01 mg/L (whichever is greater) from 0 - 4 mg/L; ± 10% from 4 - 5 mg/L
Air Purge:
Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar maximum
Alerts:
Low and high chlorine. Dirty cell. Low and high sample flow.
Certifications:
CE compliant with: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
South Korea KC Certificate
North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003
Controller Compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Cycle Time:
2.5 minutes
Dimensions (H x W x D):
342 mm x 329 mm x 177 mm
Drain:
½-inch ID flexible hose
Enclosure Rating:
IP66
Inlet:
¼-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure:
0.3 - 5.2 bar supplied to Y-strainer
0.1 - 0.3 bar supplied to analyser
Installation Kit:
Standpipe Installation Kit
Interferences:
Other oxidising agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃.
Light source:
LED, measurement at 510 nm, 1 cm light pathlength
Lower Limit of Detection (LOD):
0.008 mg/L Cl₂ at 25 °C
Mounting:
Wall mount
Operating Humidity:
0 - 90% relative humidity, non-condensing
Operating temperature range:
5 - 40 °C
Parameter:
Chlorine, total residual
Power Requirements (Voltage):
12 VDC, 400 mA maximum (supplied by the controller)
Range:
0.008 - 5 mg/L Cl₂
Reagent Consumption:
0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
Sample Flow Rate:
60 - 200 mL/min through the instrument
Sample requirements:
Y-strainer filtration with 40-mesh screen or higher
Sample Temperature:
5 - 40 °C
Warranty:
12 months
Weight:
4.1 kg
What's included?:
ULR CL17sc Chlorine Analyser, Standpipe Installation Kit, and User Manual. User provides screws and/or anchors for mounting.