Ambient Temperature:
5 - 40 °C
Dimensions (H x W x D):
2010 mm x 390 mm x 160 mm
Housing Material:
PLC Enclosure: Polystyrene with glazed polycarbonate door
Instrument air:
Instrument air: 6 bar
(50 L/min flow rate when venturi pump is running at 6 bar)
(~100 L/min flow rate when venturi sampler is purging at 3.5 bar)
Air quality: -20 °C dew point (free of water, oil and dust)
Material:
Sample chamber: PFA
Fittings and valves: SS-316, EPDM, PTFE
Venturi pump: SS-316
Sample tubing: PFA (1/2" OD x 3/8" ID)
Number of sample streams:
3
Operating System:
PLC with Keyboard
Automatic 'no sample fill' fault indication
Direct control from the BioTector through the 6 core cable
Power Consumption:
24 W (1 Amp 24 V DC), power supplied by BioTector analyser
Requirement rinse water pressure:
3 bar (~10 L/h at 3 bar)
Water can be heated up to 90 °C for hot water backwash
Water regulator and/or water heater to be supplied by the user
Sample Inlet Temperature:
2 - 90 °C
Sample Pressure:
Ambient pressure
Sample Quality:
Only for non-foaming samples!
Sampling Principle:
Vacuum sampling
Service Interval:
6 months (typical)
Suction Height:
7 m (at sea level; measured from the bottom of the sampler to the lowest liquid level)
Suction Hose Length:
Max. distance of sample transport: 80 m
Weight:
20 - 30 kg
What's included?:
Sampler only.
Hoses need to be ordered separately.